Hydromer is ISO 9001:2015 certified with TUV Rheinland of North America. ISO 13485:2016 is certified with BSI. 

FDA registered. Hydromer is ISO 9001:2015 certified with TUV Rheinland of North America. ISO 13485:2016 is certified with BSI.

Hydrophilic Coatings for Polycarbonate Substrates in Medical Devices

Table of Contents

Hydrophilic medical device coatings attract water and spread it into a thin, even film on the surface. On polycarbonate, this film keeps optical windows clear of fog and helps fluids spread evenly inside diagnostic cartridges. 

Polycarbonate is a transparent, amorphous thermoplastic that is widely used to make rigid medical device components. It combines glass-like clarity with high impact resistance, thermal stability, and dimensional stability.1 Medical device coatings can be applied to these polycarbonate components to add surface functions that the bare plastic does not offer.

Hydrophilic coatings for medical devices attract water and spread it into a thin, even film on the surface. On polycarbonate, this film keeps optical windows clear of fog and helps fluids spread evenly inside diagnostic cartridges. That said, polycarbonate is sensitive to certain solvents and can crack under stress, so the coating must be matched to it.2 At Hydromer®, our UV-curable hydrophilic coatings are suitable for polycarbonate and cure through UV-A light at a nearly ambient temperature.

Polycarbonate is one of several suitable substrates for hydrophilic medical device coatings, alongside metals, PEBAX, and glass. This article covers why polycarbonate is a preferred device material and where it is used in medical devices. We have also explained the coating challenges specific to polycarbonate and how a well-matched hydrophilic coating solves them. The goal is to help biomedical engineers, medical device manufacturers, and procurement professionals choose the right coating for their polycarbonate parts.

Key Takeaways: 

  • Hydrophilic coatings for medical devices attract water and spread it into a thin, even film on the surface.
  • A suitable coating works within polycarbonate’s limits on solvents, stress, and heat.
  • At Hydromer®, our UV-curable hydrophilic coatings are suitable for polycarbonate and cure through UV-A light at a nearly ambient temperature.
  • Adhesion testing, wettability checks, and craze inspection should run on finished, sterilized parts, not flat coupons.

The Use Of Polycarbonate As a Substrate In Medical Device Manufacturing

Polycarbonate has been used for rigid medical device parts for decades. Device makers choose it because it brings clarity, toughness, and stable dimensions together in one moldable plastic.1 Below are the key properties that make it a preferred choice:

  • Optical Transparency: Clear grades let clinicians see blood, fluids, and tissue through the part.1 This visibility is critical in clinical and diagnostic settings, where staff need to confirm flow at a glance.
  • Impact Resistance: Polycarbonate combines strength, rigidity, and toughness in one material.1 This means parts can take drops, knocks, and internal pressure without shattering.
  • Thermal Stability: It keeps its strength and shape well above room and body temperature.1 Parts hold up through warm storage and shipping without softening or warping.
  • Dimensional Stability: Polycarbonate absorbs very little water and holds its molded dimensions.1 This helps pressurized parts, such as angioplasty inflation devices, keep their shape and form tight connections.

Understanding Polycarbonate As A Coating Substrate

Polycarbonate is an amorphous polymer, which means its chains are not packed into ordered crystals. This glassy structure is what lets light pass through the part so cleanly. Its untreated surface carries relatively few polar groups, so water tends to bead up on it instead of spreading out.3

Most medical parts are injection molded from a specific grade, and medical grades differ in their additives and performance. Gamma-stabilized, lipid-resistant, and higher-heat grades are all available.1 As such, the exact grade should be named in the coating specification, along with the part design.

What Are Hydrophilic Medical Coatings?

Hydrophilic medical coatings are thin, water-loving polymer layers bonded to a device surface. Polyvinylpyrrolidone (PVP) is among the most common polymers used to make them.

In the finished coating, these polymers are held in a crosslinked network, which forms during heat or UV curing.4 The coating stays dry and easy to handle until it meets water or body fluids. It then absorbs water and swells into a soft, hydrated layer.4 Depending on the formulation, this layer makes the surface slippery, easy to wet, or resistant to fogging.

The Use Of Hydrophilic Coatings For Polycarbonate Substrates In MedTech

Hydrophilic coatings add value to polycarbonate parts where fluids, blood, or moisture meet the surface. Below are some of the medical applications where coated polycarbonate is used:

  • Diagnostic Device Housings: Polycarbonate is used for the molded housings and cartridges of diagnostic devices, including microfluidic chips.5 Sample flow in their narrow channels needs a wettable surface, and most plastics used for these chips are only mildly wettable on their own.6 A hydrophilic coating adds this wettability, so the sample moves through by capillary action.
  • Connectors And Fluid Path Components: Luer fittings, stopcocks, Y-injection sites, and check valves are often made from polycarbonate.1 On these parts, a hydrophilic layer can help fluids wet the inner walls evenly during priming and flushing.
  • Optical And Imaging Components: Clear polycarbonate windows, lenses, and viewing ports lose visibility when moisture condenses on them. Anti-fog coatings spread that condensation into a thin, clear film instead of light-scattering droplets.7
  • Blood-Handling Housings: Dialyzer, oxygenator, and blood filter housings are classic polycarbonate parts.1 Clotting on blood-contacting surfaces is a common cause of device failure.8 A hydrophilic layer binds water at the surface, which can help reduce protein and cell adhesion.9

Challenges Associated With Coating Polycarbonate Substrates

Coating polycarbonate comes with a few challenges that have to be planned for from the start. Below are the main ones:

Hydrophilic coatings for polycarbonate substrates on a clear diagnostic cartridge

1. Solvent Sensitivity

Many hydrophilic coatings are carried in organic solvents, and polycarbonate reacts badly to a lot of them. In one study, even isopropyl alcohol and methanol sharply cut the strength of polycarbonate under strain.2

A glycol ether in the same study broke a strained sample in about 22 seconds.2 A carrier that is safe on metal can still damage polycarbonate.

2. Stress Cracking And Crazing

Molded parts carry internal stress from injection molding and assembly. When a chemical reaches a stressed area, cracks can form well below the load the part normally handles.2

On clear parts, crazes appear as fine lines that cloud the view. On connectors, they can grow into cracks and leaks, as happens when lipid emulsions attack stressed Luer fittings.1

3. Limited Coating Adhesion

A hydrophilic coating swells as it absorbs water.4 That swelling pulls on the bond between the coating and the substrate. The low-polarity surface of untreated polycarbonate gives the coating few sites to anchor to.3

A weak bond lets the coating lift, peel, or shed particles once the device is in use.

Learn more about adhesion testing for hydrophilic medical device coatings. 

4. Heat Distortion During Curing

Many hydrophilic coatings are cured with heat, but standard polycarbonate grades can distort above about 121°C (250°F).1 An aggressive cure can warp the part and spoil tight features such as Luer tapers.

UV-A curing avoids this, as it cures the coating at a nearly ambient temperature and reduces heat distortion of the substrate.

5. Sterilization Compatibility

Polycarbonate handles ethylene oxide and radiation sterilization, although gamma exposure can yellow clear grades, mainly those without stabilizers.1 Repeated autoclaving does not suit it either.1

Sterilization affects the coating as well. UV-cured coatings, for example, have to keep their adhesion and wettability through the same cycle as the part.

Learn more about how UV cured coatings hold up in medical device sterilization.

Selecting A Suitable Hydrophilic Coating For Polycarbonate Substrates

A suitable coating works within polycarbonate’s limits on solvents, stress, and heat. Below are the key considerations:

  • Carrier Solvent: Choose a water-based or polycarbonate-safe carrier, and screen it on strained samples of your actual grade.2
  • Cure Method: UV and thermal coatings cure differently, so match the cure to the part’s heat limit. Note that with UV-A, the light must reach every coated surface.
  • Surface Cleaning And Activation: Surface preparation starts with removing any mold release and handling residues. Plasma activation then adds polar groups that help the coating bond.3
  • Film Thickness And Clarity: Keep the film thin and even, so it does not haze a window or change a Luer fit.
  • Validation On The Molded Part: Adhesion testing, wettability checks, and craze inspection should run on finished, sterilized parts, not flat coupons.

Hydromer® UV-Curable Coatings For Polycarbonate Substrates

The right coating for polycarbonate protects what makes the plastic worth using, such as its clarity, toughness, and tight tolerances. At Hydromer®, Inc., we have formulated hydrophilic medical coatings for over four decades. For polycarbonate parts, we offer our UV-curable hydrophilic coatings.

Our FDA-registered, ISO 13485:2016 and ISO 9001:2015 certified facility backs this with contract coating, specialized analytical testing, and in-house ISO 10993 biocompatibility testing as per your project. We can coat the parts for you, or transfer the process to your own line.

Contact Hydromer® to discuss UV-curable hydrophilic coatings for your polycarbonate parts

FAQ

Why is polycarbonate used in medical devices?

Device makers choose it because it brings clarity, toughness, and stable dimensions together in one moldable plastic.

What are hydrophilic medical coatings?

Hydrophilic medical coatings are thin, water-loving polymer layers bonded to a device surface. Polyvinylpyrrolidone (PVP) is among the most common polymers used to make them.

Why is polycarbonate hard to coat?

A hydrophilic coating swells as it absorbs water. That swelling pulls on the bond between the coating and the substrate. The low-polarity surface of untreated polycarbonate gives the coating few sites to anchor to.

Why does solvent choice matter when coating polycarbonate?

Many hydrophilic coatings are carried in organic solvents, and polycarbonate reacts badly to a lot of them. … A carrier that is safe on metal can still damage polycarbonate.

Why use UV-A curing for polycarbonate?

Many hydrophilic coatings are cured with heat, but standard polycarbonate grades can distort above about 121°C (250°F). An aggressive cure can warp the part and spoil tight features such as Luer tapers. UV-A curing avoids this, as it cures the coating at a nearly ambient temperature and reduces heat distortion of the substrate.

Editorial & Technical Review Board


To ensure the highest standards of engineering precision and scientific accuracy, this article was reviewed, validated, and approved by:

References

View All References (9 Scientific and Technical Sources)

1. Powell, D.G., Medical Applications of Polycarbonate, Medical Plastics and Biomaterials Magazine, September 1998. https://www.mddionline.com/product-development/medical-applications-of-polycarbonate

2. Al-Saidi, L.F., Mortensen, K., Almdal, K., Environmental stress cracking resistance. Behaviour of polycarbonate in different chemicals by determination of the time-dependence of stress at constant strains, Polymer Degradation and Stability, 2003, 82(3), 451-461. DOI: 10.1016/S0141-3910(03)00199-X. https://doi.org/10.1016/S0141-3910(03)00199-X

3. Kelar, J., Shekargoftar, M., Krumpolec, R., Homola, T., Activation of polycarbonate (PC) surfaces by atmospheric pressure plasma in ambient air, Polymer Testing, 2018, 67, 428-434. DOI: 10.1016/j.polymertesting.2018.03.027. https://doi.org/10.1016/j.polymertesting.2018.03.027

4. LaPorte, R.J., Hydrophilic Polymer Coatings for Medical Devices, CRC Press, 1997 (eBook edition 2017). DOI: 10.1201/9780203751381. https://doi.org/10.1201/9780203751381

5. Soper, S.A., Henry, A.C., Vaidya, B., Galloway, M., Wabuyele, M., McCarley, R.L., Surface modification of polymer-based microfluidic devices, Analytica Chimica Acta, 2002, 470(1), 87-99. DOI: 10.1016/S0003-2670(02)00356-2. https://doi.org/10.1016/S0003-2670(02)00356-2

6. Olanrewaju, A., Beaugrand, M., Yafia, M., Juncker, D., Capillary microfluidics in microchannels: from microfluidic networks to capillaric circuits, Lab on a Chip, 2018, 18(16), 2323-2347. DOI: 10.1039/C8LC00458G. https://doi.org/10.1039/C8LC00458G

7. Chu, J., Tian, G., Feng, X., Recent advances in prevailing antifogging surfaces: structures, materials, durability, and beyond, Nanoscale, 2023, 15(27), 11366-11402. DOI: 10.1039/D3NR01767B. https://doi.org/10.1039/D3NR01767B

8. Jaffer, I.H., Fredenburgh, J.C., Hirsh, J., Weitz, J.I., Medical device-induced thrombosis: what causes it and how can we prevent it?, Journal of Thrombosis and Haemostasis, 2015, 13(Suppl 1), S72-S81. DOI: 10.1111/jth.12961. https://doi.org/10.1111/jth.12961

9. Chen, S., Li, L., Zhao, C., Zheng, J., Surface hydration: Principles and applications toward low-fouling/nonfouling biomaterials, Polymer, 2010, 51(23), 5283-5293. DOI: 10.1016/j.polymer.2010.08.022. https://doi.org/10.1016/j.polymer.2010.08.022