Hydrophilic Coatings for Nitinol Substrates: A Comprehensive Guide for Medical Devices

Table of Contents

Hydrophilic coatings for nitinol substrates absorb water and cut friction between nitinol parts and the vessel wall.

Nitinol is a near-equiatomic alloy of nickel and titanium, best known for its superelasticity and shape memory.1 Both behaviors come from a reversible phase shift, which lets the alloy recover strains of up to about 10%.1, 2 Medical device coatings on nitinol parts must follow this movement without cracking or lifting. Suitable substrates for hydrophilic medical device coatings include several metals, but few of them spring back like nitinol.2

Nowadays, nitinol is found in stents, heart valves, guidewires, and vena cava filters, and its use keeps growing.1 Hydrophilic coatings for medical devices absorb water and cut friction between nitinol parts and the vessel wall.3 When such a medical coating cannot keep up with the alloy, it peels and sheds into the bloodstream.4

In this article, we will discuss how nitinol behaves as a substrate and how hydrophilic coatings bond to its surface. Our focus will be on understanding the benefits of coating nitinol and the devices that use it. Next come the adhesion challenges tied to its shape memory and the selection checks that help ensure a lasting bond.

Key Takeaways

  • Coating nitinol pairs the alloy’s mechanics with a wet, low-friction surface. The hydrated layer lets a nitinol wire slide along the vessel wall with less drag.
  • The coating-to-nitinol bond forms at the oxide surface, so surface preparation decides much of its strength.
  • Nitinol’s shape-memory behavior is what makes it useful, and it is also what makes coating adhesion difficult.
  • Coating selection for nitinol comes down to strain, surface, heat, and test evidence.

Understanding Nitinol As a Substrate for Hydrophilic Coatings

Nitinol is superelastic above its austenite finish temperature (Af) and easy to deform below it.1, 2 Self-expanding stents typically have an Af near 30°C (86°F), so they crimp at room temperature and spring open in the body.2 Nitinol’s surface carries a titanium oxide layer that forms on its own.2 Both the thickness and the nickel content of this oxide change with processing.5

Shape-setting heat treatments can thicken the oxide many times over.6 Electropolishing removes this heat-treatment oxide and leaves a uniform, titanium-rich layer.2, 5 As such, the coating specification should name the surface finish, since the coating bonds to that oxide layer. Hydrophilic coatings for metal substrates deal with the same oxide question on other device metals.

What Are Hydrophilic Medical Coatings?

A hydrophilic medical coating is a polymer film that draws water to a device surface and holds it there. It is built from polymers that carry water-absorbing groups, such as polyvinylpyrrolidone (PVP), polyethylene glycol (PEG), polyvinyl alcohol, and polysaccharides.3 After contact with water or saline, the coating takes up water and forms a lubricating layer.3, 7

This layer reduces friction between a device and the vessel wall.7 We call this behavior “slippery when wet.”

How Hydrophilic Coatings Interact with Nitinol Surfaces

The coating-to-nitinol bond forms at the oxide surface, so surface preparation decides much of its strength. In one wire study, etched and pickled surfaces held a polymer coating far better than mechanically polished ones.6 In another, surface modification of a nitinol stent raised the peel strength of a polymer coating by 300%.8 This is why cleaning and pretreatment matter for every nitinol part.

Benefits of Applying Hydrophilic Coatings to Nitinol Substrates

Coating nitinol pairs the alloy’s mechanics with a wet, low-friction surface. The alloy still does the bending and recovering, while the coating handles contact with blood and the vessel wall. This pairing pays off in four ways:

  • Lower Friction Against the Vessel Wall: The hydrated layer lets a nitinol wire slide along the vessel wall with less drag.7 With nitinol’s kink resistance on top, the wire can follow tight bends without buckling.9
  • Less Platelet Adhesion: In lab tests on nitinol plates, a hydrophilic polymer coating cut platelet coverage from about 49% to about 1%.10 This matters on stents and other parts that stay in contact with blood.
  • Nanometer-Scale Thickness: Hydrophilic coatings under 12 nm thick have been bonded covalently to braided nitinol-based wires.11 Layers this thin add almost nothing to the profile of fine struts and braids.
  • A Normal Healing Response: In a 30-day rabbit study, coated and uncoated nitinol flow diverters showed similar inflammation.11 Both the coated and the uncoated devices were fully covered by endothelium. This suggests the coating adds no extra inflammatory response under normal operating conditions.

Medical Devices That Use Hydrophilic-Coated Nitinol

Nitinol is used across vascular and minimally invasive devices.1 Its uses range from vena cava filters and septal occluders to heart valve frames and stone retrieval baskets.1, 9 Three device groups show most clearly where a hydrophilic layer on nitinol earns its place:

  • Guidewires: Angioplasty guidewires were among the first devices to use nitinol’s kink resistance.9 Hydrophilic guidewire coatings need to stay bonded, as guidewires were among the devices in the coating separation recalls the FDA reported in 2015.4
  • Self-Expanding Stents: Laser-cut tubing is the most preferred choice for these stents.2, 9 They are made larger than the target vessel and open on their own once the delivery sheath is pulled back.2 Their struts stay in contact with blood for the life of the implant.
  • Flow Diverters: These neurovascular stents are braided from fine nitinol-based wires.10, 11 Hydrophilic polymer layers on these wires showed minimal platelet adherence in lab tests.10

Challenges Associated with Coating Nitinol Substrates

Nitinol’s shape-memory behavior is what makes it useful, and it is also what makes coating adhesion difficult. Five issues stand out:

ChallengeEffect on the CoatingReported Data
Superelastic strain during crimpingA coating that cannot follow the strain can crack or lift off the strutsStrains of up to 8% while held in the delivery system5
Non-superelastic surface oxideA coating anchored to the oxide can crack and lift along with itA very thick oxide cracked at 3% bending strain (one study)5
Heat-sensitive transformation temperaturesAny heated step after shape setting, including a thermal cure, adds to the thermal historyAf shifted most between about 350°C and 450°C (662–842°F) and very little at 300°C (572°F)12
Nickel release and corrosionPretreatment must not leave a surface that corrodes fasterThicker thermal oxides cut the corrosion breakdown potential sharply (one study)5
Fatigue under cyclic loadingBoth the stent and its coating have to survive every cycleAbout 40 million pulsatile cycles a year; fatigue tests run to 400 million cycles13
  1. Superelastic Strain During Crimping: A self-expanding stent can see strains of up to 8% while held in its delivery system.5 A coating that cannot follow that strain can crack or lift off the struts.
  2. A Surface Oxide That Is Not Superelastic: The oxide on nitinol does not share the alloy’s superelasticity.5 In one study, a very thick oxide cracked badly at 3% bending strain.5 A coating anchored to that oxide can crack and lift along with it.
  3. Heat-Sensitive Transformation Temperatures: Nitinol’s transformation temperatures shift with small changes in process temperature or heating time.1 In one aging study on medical wire, Af shifted most between about 350°C and 450°C (662–842°F) and very little at 300°C (572°F).12 Any heated step after shape setting, including a thermal cure, adds to this thermal history.
  4. Nickel Release and Corrosion: Nitinol’s corrosion resistance depends heavily on surface processing, and corrosion can release nickel ions.1 In one study, thicker thermal oxides cut the corrosion breakdown potential sharply.5 Pretreatment for adhesion must not leave a surface that corrodes faster.
  5. Fatigue Under Cyclic Loading: A stent sees about 40 million pulsatile cycles a year, and fatigue tests run to 400 million cycles.13 Peripheral stents also bend, twist, and shorten with leg movement.13 Both the stent and its coating have to survive every one of these cycles.

When adhesion fails, a coating can peel, flake, or shed from the device. FDA has received reports linking such separation to embolism, stroke, and death.4

Selecting a Suitable Hydrophilic Coating for Nitinol Substrates

Coating selection for nitinol comes down to strain, surface, heat, and test evidence. The checks below cover each of them:

  • Coating Flexibility: Choose a coating that bends and recovers with the alloy. Prove it on parts that have been crimped, loaded, and deployed.
  • Oxide Control: Surface preparation for nitinol should strip thick heat-treatment oxide and then passivate the surface.5
  • Cure Profile: Log the cure temperature and time, and measure Af again on coated parts.1 This helps ensure the part still behaves as designed at body temperature.
  • Film Build: Aim for a thin, uniform layer that coats each strut and wire without adding bulk.
  • Particulate Testing: FDA’s guidewire guidance asks for particle counts after simulated use, reported at ≥10 μm, ≥25 μm, and ≥50 μm.14

Corrosion Testing: Note that for polymer-coated nitinol, FDA points to an uncoated part with identical thermal processing as the worst case for corrosion testing.1

Hydromer®: Thermal-Curable Hydrophilic Coatings and Coating Services for Nitinol

Hydromer®, Inc. brings over 45 years of coating experience to customers across the United States, Europe, and Asia-Pacific. For nitinol parts, our thermal-curable hydrophilic coatings cover NiTi alloys, complex geometries, and internal lumens. Our coatings can be applied through dip, proprietary meniscus, flow, spin, spray, micro-application, or internal lumen (ID) coating, as per your project.

5 Top Advantages of Choosing Hydromer® Coatings for Nitinol Devices

Here is what a nitinol program gains with us:

  • Certified Quality System: Our FDA-registered, ISO 13485:2016 and ISO 9001:2015 certified facility follows FDA GMP (21 CFR 820).
  • Particle Data in FDA Size Bands: Particle counts for our thermal-curable coatings are reported at ≥10 μm, ≥25 μm, and ≥50 μm.14
  • In-House Testing: Our Ph.D. chemists run coefficient of friction (COF), contact angle, cytotoxicity, and endotoxin tests. ISO 10993 biocompatibility testing is also done in-house.
  • Regulatory Support: Nearly all of our coatings are covered by Master Access Files (MAFs), and our regulatory team supports 510(k) and PMA submissions.
  • Prototype to Production: Contract coating covers prototypes through market-ready devices, and technology transfer adds custom coating machines with on-site training.

Let us review your nitinol part, its surface finish, and its Af window with you. Ask for a free sample, or set up a call with a coatings engineer.

Frequently Asked Questions

Why coat nitinol with a hydrophilic coating?

Coating nitinol pairs the alloy’s mechanics with a wet, low-friction surface. The alloy still does the bending and recovering, while the coating handles contact with blood and the vessel wall.

Why does the surface oxide matter for hydrophilic coatings on nitinol?

Shape-setting heat treatments can thicken the oxide many times over. Electropolishing removes this heat-treatment oxide and leaves a uniform, titanium-rich layer. As such, the coating specification should name the surface finish, since the coating bonds to that oxide layer.

What happens to a hydrophilic coating when a nitinol stent is crimped?

A self-expanding stent can see strains of up to 8% while held in its delivery system. A coating that cannot follow that strain can crack or lift off the struts.

Which devices use nitinol?

Nitinol is used across vascular and minimally invasive devices. Its uses range from vena cava filters and septal occluders to heart valve frames and stone retrieval baskets.

How does heat curing affect hydrophilic coatings for nitinol substrates?

Nitinol’s transformation temperatures shift with small changes in process temperature or heating time. In one aging study on medical wire, Af shifted most between about 350°C and 450°C (662–842°F) and very little at 300°C (572°F). Any heated step after shape setting, including a thermal cure, adds to this thermal history.

Editorial & Technical Review Board


To ensure the highest standards of engineering precision and scientific accuracy, this article was reviewed, validated, and approved by:

References

View All References (14 Scientific and Regulatory Sources)
  1. U.S. Food and Drug Administration, Technical Considerations for Non-Clinical Assessment of Medical Devices Containing Nitinol, Guidance for Industry and FDA Staff, July 9, 2021. https://www.fda.gov/media/123272/download
  2. Stoeckel, D., Pelton, A., Duerig, T., Self-expanding nitinol stents: material and design considerations, European Radiology, 2004, 14(2), 292-301. DOI: 10.1007/s00330-003-2022-5. https://doi.org/10.1007/s00330-003-2022-5
  3. Niemczyk, A., El Fray, M., Franklin, S.E., Friction behaviour of hydrophilic lubricious coatings for medical device applications, Tribology International, 2015, 89, 54-61. DOI: 10.1016/j.triboint.2015.02.003. https://doi.org/10.1016/j.triboint.2015.02.003
  4. U.S. Food and Drug Administration, Lubricious Coating Separation from Intravascular Medical Devices: FDA Safety Communication, November 23, 2015. http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm473794.htm
  5. Zhu, L., Trépanier, C., Fino, J., Pelton, A.R., Oxidation of Nitinol and its Effect on Corrosion Resistance, Proceedings of the ASM Materials & Processes for Medical Devices Conference, ASM International, 2004, 156-161. https://confluentmedical.com/wp-content/uploads/2023/05/162_Zhu_Trepanier_Fino_Pelton_2004.pdf
  6. Carroll, W.M., Rochev, Y., Clarke, B., Burke, M., Bradley, D.J., Plumley, D.L., Proft, J., Influence of Nitinol Wire Surface Preparation Procedures, on Cell Surface Interactions and Polymer Coating Adherence, Proceedings of the Materials and Processes for Medical Devices Conference, Anaheim, CA, 2003. https://www.fwmetals.com/media/bwphynko/influence-of-nitinol-wire-surface-preparation-procedures-on-cell-surface-interactions-and-polymer-coating-adherence.pdf
  7. U.S. Food and Drug Administration, Intravascular Catheters, Wires, and Delivery Systems with Lubricious Coatings – Labeling Considerations, Guidance for Industry and FDA Staff, October 10, 2019. https://www.fda.gov/media/113951/download
  8. Bakhshi, R., Darbyshire, A., Evans, J.E., You, Z., Lu, J., Seifalian, A.M., Polymeric coating of surface modified nitinol stent with POSS-nanocomposite polymer, Colloids and Surfaces B: Biointerfaces, 2011, 86(1), 93-105. DOI: 10.1016/j.colsurfb.2011.03.024. https://doi.org/10.1016/j.colsurfb.2011.03.024
  9. Duerig, T., Pelton, A., Stöckel, D., An overview of nitinol medical applications, Materials Science and Engineering A, 1999, 273-275, 149-160. DOI: 10.1016/S0921-5093(99)00294-4. https://doi.org/10.1016/S0921-5093(99)00294-4
  10. Lenz-Habijan, T., Bhogal, P., Peters, M., Bufe, A., Martinez Moreno, R., Bannewitz, C., Monstadt, H., Henkes, H., Hydrophilic Stent Coating Inhibits Platelet Adhesion on Stent Surfaces: Initial Results In Vitro, CardioVascular and Interventional Radiology, 2018, 41(11), 1779-1785. DOI: 10.1007/s00270-018-2036-7. https://doi.org/10.1007/s00270-018-2036-7
  11. Lenz-Habijan, T., Bhogal, P., Bannewitz, C., Hannes, R., Monstadt, H., Simgen, A., Mühl-Benninghaus, R., Reith, W., Henkes, H., Prospective study to assess the tissue response to HPC-coated p48 flow diverter stents compared to uncoated devices in the rabbit carotid artery model, European Radiology Experimental, 2019, 3, 47. DOI: 10.1186/s41747-019-0128-z. https://doi.org/10.1186/s41747-019-0128-z
  12. Pelton, A.R., DiCello, J., Miyazaki, S., Optimization of Processing and Properties of Medical-Grade Nitinol Wire, Proceedings of the International Conference on Shape Memory and Superelastic Technologies (SMST-2000), 2000, 361-374. https://confluentmedical.com/wp-content/uploads/2023/05/027.pdf
  13. Pelton, A.R., Schroeder, V., Mitchell, M.R., Gong, X.-Y., Barney, M., Robertson, S.W., Fatigue and durability of Nitinol stents, Journal of the Mechanical Behavior of Biomedical Materials, 2008, 1, 153-164. DOI: 10.1016/j.jmbbm.2007.08.001. https://doi.org/10.1016/j.jmbbm.2007.08.001
  14. U.S. Food and Drug Administration, Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling, Guidance for Industry and FDA Staff, October 10, 2019. https://www.fda.gov/regulatory-information/search-fda-guidance-documents/coronary-peripheral-and-neurovascular-guidewires-performance-tests-and-recommended-labeling